Calibration
































The Calibration Tab is used to set the Zero (LRV) and Span (URV) points for the Level Float (Level #1), Interface Float in
a 2 float system (Level #2), and the Temperature sensor (if installed). To change a value, click the cell you want to
change, type in the new value, and click Write.

The Units field shows the unit of measure used for the Zero and Span points. To change, click the drop-down arrow, and
select the desired unit of measure. Click “Write” when finished. Make sure that the Zero and Span values match the unit
of measure selected. This program does not convert between different units.

The Offset field is used for making adjustments to the optional display, not the output current. For example, if the float
(Level #1) is at the 4mA point, but the display reads 0.25”, enter -0.25 in the Offset field and click “W” (Write). The display
will now read 0.00” even though the current has remained at 4mA.