Setting equilibration time – VICI FTP User Manual

Page 12

Advertising
background image

8

Programming and Running a Sequence

3. Type SS2<enter> and type the values below for example state 2, followed by <enter>.

: ss2

Active (Y)? (no change)

Ramp rate, degC/min (0):100

Setpoint (0):100

Hold time, seconds (0):10

3. Type SS3<enter> and type the values below for example state 3, followed by <enter>.

: ss3

Active (N)? Y

Ramp rate, degC/min (0):200

Setpoint (0):200

Hold time, seconds (0):30

4. Type SS4<enter> and type the values below for example state 4, followed by <enter>.

: ss4

Active (N)? Y

Ramp rate, degC/min (0):200

Setpoint (0):300

Hold time, seconds (0):25

Always verify that changes were made by repeating the command
to view the current state, or SS<enter> to show all states.

5. Type SS<enter>. If the data were entered correctly, the screen will reflect the following

information (except for the italicized explanatory text).

State Active Rate Temp Time

1

Yes

0

25

5

Start at 25°C, hold for 5 seconds (no ramp in state 1)

2

Yes

100

100

10

Ramp @ 100°C/min to 100°C, hold 10 seconds

3

Yes

200

200

30

Ramp @ 200°C/min to 200°C, hold 30 seconds

4

Yes

200

300

25

Ramp @ 200°C/min to 300°C, hold 25 seconds

5

No

0

0

0

If no other state is active, the fan will automati-
cally come on to cool the column down to the
State 1 temperature. The FTC will hold “Ready”
for the next run. (Refer to the section entitled “Fan
Modes” on page 9.)

6

No

0

0

0

7

No

0

0

0

8

No

0

0

0

NOTE: State parameters are saved when the FTC is powered down. Pushing the RESET
button on the front of the FTC may erase the state parameters.

Setting Equilibration Time

Equilibration Time delays the “Ready” flag for state 1, allowing time for the column
temperature to equalize once the temperature has been reached. A “Go” command will
not execute until this time has elapsed. The equilibration time is preset to 5 seconds. To
change it:

1. Type ET=n<enter>, where n is the desired equilibration time, in seconds.

2. Type ET<enter> to verify that the change has been executed.