Viewing the command set – VICI FTP User Manual
Page 9

5
Getting Started
Viewing the Command Set
To view the list of commands for the FTC, type ??<enter>. All HyperTerminal commands
must be followed by <enter> on the keyboard. Characters may be in either upper or
lower case. You may wish to print the list for for reference.
A+
Enable autocorrection
A-
Disable autocorrection
CC
Connect column
CD
Disconnect column
CI
Calibrate current scale
CR
Calibrate column resistance
CV
Calibrate voltage scale
D+
Debug on
D-
Debug off
DD
Dump data
DR
Dump ram
DX
Dump parameters to Xml
E+
Terminal echo on
E-
Terminal echo off
EC
Enter R(t) calibration data
EM
Enter MPC parameters: [offset [skip]]
ES
Enter state data
ET[=n] Show/set equilibration time (seconds)
F+
Fan on
F-
Fan off
FM[=n] Show/set fan mode: [0]off, [1]on, [2]temp, [3]time
FT[=n] Show/set fan temp (°C) / time (seconds)
GO
Start ramp/cooling cycle
L+
Start data logging
L-
Stop data logging
LA[=n] Show/set matrix lambda
RR
Read column resistance
RS
Reset state machine
RT
Read column temperature
RX
Read parameters from Xml. Use flow control!
SR
Run step response test
SS[n]
Show/set state (n=1-8) parameters
VR
Show firmware version
WP
Write eprom parameters
Z-
Turn off column heating
Z+
Turn on column heating
??
Show this list