Getting Started

6. Click Apply, then OK.

7. On the next screen, select File > Properties >

Settings, and click on ASCII Setup.

8. In the ASCII setup window, click the check

boxes or enter the values as required to
match the screen in Figure 3.

9. Click OK, then OK again to close the FTC

Properties window.

10. HyperTerminal:

Click on the phone icon just below “View”
to connect. If the connection is not
available, repeat the procedure, selecting
a different COM port in step 4.

11. When HyperTerminal is connected, press the

reset button of the FTC. The FTC will read as
follows:

FTC Response

Explanation

Resetting PWM . . . ok

Pulse Width Modulator working

A/D started

Analog to Digital converter working

EEPROM parameters recovered

Operational parameters loaded

DS24B33 present

Memory chip recognized

1-wire checksum ok

Verify chip is OK

ID = 00000142F5DA

Chip ID number

Reading embedded parameters

Reading column data

Column id : 12345

Column serial number

Length m : 10

Column length in meters

Phase id : VB-1 250um

Column phase, film thickness (µm) & inside diameter (µm)

Element equation :

Column temperature vs. resistance equation

R(T) = 1.31039E1 + 6.44364E-2 * T + 1.74357E – 4 * T^2

M = 1.550

Heat capacity of the column

Alpha = 2.10000E-2

Time constant in seconds, how fast does the temp. rise

Lambda = 1.00000E-4

PWM dampener

Max. temp = 340

Column maximum temperature

Calibration data :

Column resistance and temperature calibration data

1 40.000 5.600

pair #1

(*C)

(ohms)

2 183.000 10.692

pair #2

(*C)

(ohms)

3 300.000 17.785

pair #3

(*C)

(ohms)

OK

Data accepted

Weighting vector generated.

Algorithm calculation complete

Checking input connections . . . ok

If not “ok”, check column connections

0.4.8.095 July 5 2012 12:30:55, Unit #321 FTC firmware and serial number
DSP Column Control online

FTC ready for use

Figure 4: ASCII setup