Logging Column Data

Data logging allows the user to view column information collected by the FTC. Turn on
data logging, type L+<enter>. To stop logging data, type L-<enter>.

The data is presented as follows:

Total Time (sec)

Target Temp

Measured Temp

Tt-Tm

Voltage

Current

PWM

State Flag

00000001.0

40.00

40.000

0.000

0.6110

0.1038

983

1R

00000002.0

40.00

40.000

0.000

0.6110

0.1038

983

1R

00000003.0

40.00

40.000

0.000

0.6110

0.1038

983

1R

00000004.0

40.00

40.000

0.000

0.6110

0.1038

983

1R

00000005.0

40.00

40.000

0.000

0.6110

0.1038

983

1R

00000006.0

40.00

40.000

0.000

0.6110

0.1038

983

1R

The first digit of the last value denotes the number of the active state. The second digit
conveys the following meaning:

Flag

Meaning

A

Ramp auto correction on

D

PWM duty cycle limit reached

F

Fan is on

I

Column current limit reached

H

Holding in state

P

Column power limit reached

R

Ready

V

Column voltage limit reached

X

External start command received

Turning the Column On

Power to the column must be turned on before the programmed states will execute. To
do this:

1. Type Z+<enter>. The following message will appear:

WARNING: Operating with incorrect parameters may damage column!

Column temperature reading: 19.7.

Is this correct (N)?

2. If the temperature is appropriate for its environment, type Y<enter> and the FTC will

power up the column. If the column temperature reading is inaccurate, the column
may be damaged. Type N<enter> and the column will not receive power.

Incorrect column data can contribute to permanent damage
to the column. If you get a suspicious reading, please contact
Valco for further assistance.

Typing Z-<enter> will turn the column off.