9

Installation

6. Press the <Config> key, then press <Back det> or <Front Det>, corre-

sponding to the location where the PDD will be installed. For this example
we will select Back Det. The display will read:

Configured (or Unconfigured)
Ignore ready = False
[B-DET] = (Signal) (FID)

7. Press <Mode/Type>. The display will read:

CONFIGURE BACK DETECTOR
Remove Module < (cursor is here)

8. Press <Enter> to clear the existing configuration. The display will read:

CAUTION: Instrument power must be turned off
and back on for set point to take effect.

9. Do

not turn off the GC; instead press <Enter>. The display will read:

CONFIGURE BACK DETECTOR
Unconfigured: < (cursor is here)
[B-Det] = (Signal) (FID)
[EPC4] Not Found
[A-Det2] Not Found

10. Press <Mode/Type>. The display will read:

CONFIGURE BACK DETECTOR
CPDET FID,

No Htr, No EPC <

(cursor is here)

CPDET FID, Htr, No EPC

MODEL D3 IN

BACK POSITION

ELECTROMETER CABLE

CONNECTIONS

HEATER CABLE

CONNECTIONS

FRONT OF 7890

FID LOGIC BOARD

FOR BACK POSITION

FID LOGIC BOARD

FOR FRONT POSITION

THUMBSCREW

B-DET CABLE

CONNECTION

F-DET CABLE

CONNECTION

F-DET

B-DET

Figure 5: 7890 FID logic board