Troubleshooting, High background current low sensitivity, No peaks – VICI D-3-I-7890 User Manual

Page 22

Advertising
background image

18

Troubleshooting

High Background Current

Low Sensitivity

Purple/

blue

Pink

Disconnect the column
and cap the column inlet

Baseline current is still high
(> 2000 pA)

Leaks in the fittings of the
PDD plumbing

No detectable leaks.

Do a hydrogen leak test
(described on next page)

Check
discharge
color

Leaks inside the detector
cell itself

Consult

Valco

Tighten leaking fittings

The detector isn’t leaking.
Check for column bleed
and/or leaks in the GC setup.

Baseline current falls to
normal range (< 2000 pA)

Possible bad carrier gas or
empty He bottle. Install a new
bottle.

Consult

Valco

Consult

Valco

Consult

Valco

Baseline current is still high
(> 2000 pA)

Baseline current falls to
normal range (< 2000 pA)

Baseline current is still high
(> 2000 pA)

Baseline current falls to
normal range (< 2000 pA)

Baseline current is still high
(> 2000 pA)

Baseline current falls to
normal range (< 2000 pA)

Sensitivity is in

acceptable range

Sensitivity is still

too low

Sensitivity is in

acceptable range

Sensitivity is still

too low

Sensitivity is in

acceptable range

Sensitivity is still

too low

Sensitivity is in

acceptable range

Remove column and

condition it, then run

a sample

Background current

remains high

Background current

drops to normal range

Check column inlet for

leaks. Tighten fittings.

Background current

increases

Check column position

according to drawing on

p. 14, and run a sample.

Background current

remains in normal range

Remove column and

bake detector at

300°C for four hours.

Install column and run

a sample.

Run

a sample

Refer to “Troubleshooting:

High Background Current”

Reinstall column and check

background current

Normal background current

(£ 2000 pA)

High background current

(> 2000 pA)

Consult

Valco

Sensitivity is still

too low

Check background current

with column removed and

column inlet plugged

No Peaks

If the background current is stable but there are no peaks:

1. Check column flow.

2. Check the column insertion position (10.3 cm for capillary columns,

8.7 cm for packed columns).

3. If there are still no peaks, check to make sure that the GC is actually making

an injection.

4. If there are still no peaks, consult Valco.