Detector performance log – VICI D-3-I-7890 User Manual
Page 25

21
Detector Performance Log
In addition to the occasions indicated in the Comments area of the table below (see
Initial Power-Up,
page 13), the standing current should be observed and logged after any system change. Logging
the standing current (with and without the column) on a regular basis is also an effective monitor of
system integrity (leaktightness and cleanliness). To check sensitivity continuity, we recommend
tracking the internal standard (quantity on column/area count). Additional log pages can be down-
loaded from the support/manuals section of www.vici.com.
Date
Detector Model:
Serial Number:
Date of purchase:
Operator
Column
temp
(°C)
Detector
temp
(°C)
Standing
current
(pA)
Noise
level
Sample
Comments
Initial power-up (IPU)
100
IPU, detector to analysis temp
IPU, column installed
Ambient
IPU, column to analysis temp
D-3-I-