VICI TCD2 User Manual

Page 20

Advertising
background image

16

Installation

The test chromatogram accompanying your TCD was obtained under the
following conditions:

Sample:

~100 ppm blend in Helium balance

Sample volume:

250 µl

Column:

10' x 1/16" x 0.040" molecular sieve 5Å micropacked

Column flow:

6 mL/min

Reference flow:

6 mL/min

Column temperature: 65°C
Detector temperature: 100°C
Filament temperature: 230°C (a setting of 5.0)
Carrier gas:

Helium

Balancing the Channels

using LCD Display

The LCD display indicates the signal level from the recorder/chart output.

10. Set the recorder switch to A, and use the coarse and fine controls for

Channel A to adjust the signal level to about 300 mv.

11. Set the recorder switch to B and repeat the same process to adjust the

signal level to about 200 mv.

12. Set the recorder switch to A-B. The display should read about 100 mv. If

the signal levels drifts below zero, readjust the A & B channels.

The detector is now ready for analytical use.

Balancing the Channels

using Recorder/Chart Output

10. Use the zero (or shunt) setting on the recorder to set the true zero on the recorder.

11. Set the Recorder switch to A, and use the coarse and fine zero controls

for Channel A to bring the recorder pen on scale and to a position
approximately 1-2 cm above zero.

12. Set the Recorder switch to B, and repeat the same process, bringing the

pen to a position

half as far above zero as the pen for Channel A.

13. Set the Recorder switch to A - B, and observe the recorder pen position.

It should be above the true zero position, but below the position for
Channel A. If the pen drifts below zero, readjust the A and B channels.

The detector is now ready for analytical use.

Balancing the Channels

using Unattenuated Outputs

10. Connect the output cable to the A connector, and use the coarse and fine

zero controls for Channel A to bring the baseline signal on scale and to a
position approximately 100 mV above zero.

11. Move the output cable to the B connector, and use the coarse and fine

zero controls for Channel B to bring the baseline to a position half as far
above zero as for Channel A (the value of A minus B).

12. Move the output cable to the A - B connector, and observe the baseline

signal. It should be above the true zero position. If the signal drifts below
zero, readjust the A and B channels.

The detector is now ready for analytical use.

Single filament mode

The basic procedures and temperature settings described for referenced
mode operation can also be applied to operation in the single filament or
independent mode. Both channels must have the same carrier gas; you
cannot use helium carrier in one channel and nitrogen in the other. Also be
aware of the possibility of detector cross-talk when components elute from
the columns in concentrations greater than 1%.

Advertising
This manual is related to the following products: