Troubleshooting, Troubleshooting chart, Heater fault determination – VICI TCD2 User Manual

Page 21

Advertising
background image

17

Troubleshooting

Troubleshooting Chart

Problem

Possible Cause

Refer to page:

No signal

Main or filament power switch off

4

Filament temperature too low

4

Fuse blown

3

Output cable in the wrong connector

14

Signal reversed

Signal wires reversed

14

Noisy signal

Dirty system or leaking fittings

7, 13, 19

Bad flow controller in system

15

Cell/filament differential too small

15

Filament damaged

2

Cannot zero the signal

Recorder/Int. not zeroed

16

Voltage incompatibility

14

Main or filament power switch off

4

Excessive drift*

Block temperature not stabilized

15

Heater fault

17

Cell doesn’t heat properly

Heater/sensor cable unplugged/

heater fail safe mechanism

17

Heater fault

17

Fuse blown

3

For technical assistance from Valco Instruments, call (713) 688-9345 or email [email protected].

*Try referenced mode, page 12

Heater Fault Determination

The heater control circuit is designed to disable itself in the event of a fault
such as over-temperature, shorted heater, short/open temperature sensor,
etc. Once disabled, circuit operation is restored by turning off the controller,
disconnecting and reconnecting the detector cable, and then turning the
controller back on.

E.g., if mains power is applied before the detector is

connected to the controller, the heater circuit will not operate until mains
power is removed, the detector connected, and mains power restored.

Under normal circumstances, the amber temperature indicator light will flash
once after mains power is applied, before control is established. If the light
fails to flash, it indicates that a fault has been detected or (less likely) the
heater supply fuse has failed.