The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.

002 Aluminum

Blank Done

* Scan Sample

Place the reacted sample in the chamber, close the
lid and press

Q/ENTER to scan sample. The result

will be displayed.

002 Aluminum

* Scan Sample

The displayed result can now be standardized. Use
the

s or t buttons to scroll to Calibrate. Press

Q/ENTER to select.

002 Aluminum

0.09 ppm

* Scan Sample

Next Test

Previous Test

%T/Abs

Calibrate

Scan Blank

A reverse font (dark background with light
characters) will appear to indicate that the reading
can be adjusted. Use

s or t to scroll to the

concentration of the sample, 0.10 ppm in this
example.

002 Alumninum

0.09

* Calibrate

Set the calibration by pressing

Q/ENTER to select

Calibrate

.

002 Aluminum

0.10

* Calibrate

SMART2 SOIL MANUAL 1.07