LaMotte SMART 2 Electronic Soil Lab User Manual
Page 62
PROCEDURE
1. Press and hold ON button until colorimeter turns on.
2. Press ENTER to start.
3. Press ENTER to select TESTING MENU.
4. Select ALL TESTS (or another sequence containing 32 Cu DDC) from
TESTING MENU
.
5. Scroll to and select 32 Cu DDC from menu.
6. Fill a clean tube (0290) to the 10 mL line with the soil extract then
neutralize according to the procedure on page 50.
7. Insert tube into chamber, close lid and select SCAN BLANK.
8. Remove tube from colorimeter and add 5 drops of *Copper 1 (6446). Cap
and mix. Solution will turn yellow if copper is present.
9. Insert tube into chamber, close lid and select SCAN SAMPLE. Multiply the
result by 5 to determine the copper concentration in ppm.
10. Press OFF button to turn colorimeter off or press EXIT button to exit to a
previous menu or make another menu selection.
NOTE: The reaction may stain the tubes. Scrub the tubes thoroughly after
each use.
n
COPPER CONCENTRATION CHART
PARTS PER MILLION
RANGE
0-1 ppm
Low
1-3 ppm
Marginal
3-4 ppm
Adequate
62
SMART2 SOIL MANUAL 1.07