ELECTRONIC TEST METHODS

pH

PROCEDURE

1. Use the 10 g Soil Measure (1164) to add one level measure of the soil

sample to a 50 mL beaker (0944). Use the graduated cylinder (0416) to add
10 mL of deionized water. Stir thoroughly.

2. Let stand for at least 30 minutes, stirring two or three times.
3. Stir mixture just prior to making the pH reading. Determine the pH

reading of the sample by following the instructions for the pH Meter.

LIME REQUIREMENT -

WOODRUFF METHOD

PROCEDURE

1. Use the 10 g Soil Measure (1164) to add one level measure of the soil

sample to a 50 mL beaker (0944). Use the graduated cylinder (0416) to add
10 mL of deionized water. Stir thoroughly.

2. Let stand for at least 15 minutes.
3. Add 20 mL of Woodruff Buffer Solution (5272). Mix well, and let stand for

at least 20 minutes, stirring two or three times.

4. Take reading using the pH meter. Stir mixture just prior to making reading.
5. Each 0.1 pH unit drop from pH 7.0 indicates a lime requirement equivalent

to 1000 lbs calcium carbonate (CaCO

3

)/acre.

SMART2 SOIL MANUAL 4.08