EPA COMPLIANCE

To qualify as an EPA accepted test, and to achieve the greatest accuracy, the
Sodium Thiosulfate Solution, 0.025N (4169) must be standardized daily. This
procedure follows Standard Methods for the Examination of Water and
Wastewater. Numbers in ( ) are for LaMotte products. These products are not
included in this kit but can be ordered from LaMotte Company by using the
specified code number.

8

4.

5.

2.

3.

1.

6.

Add 2 drops
of Sulfuric

Acid, 5N

(8517WT).

Use a 10 mL graduated
cylinder (0416) to add

15 mL of Deionized

Water (5115) to the

titration tube (0608).

Use a Direct
Reading Titrator,

0-1 Range (1.0

mL capacity)

(0376) to add 2
mL of Potassium

Bi-iodate (7346).

Use the 0.1 g
spoon (0699) to

add 0.2 g

Potassium Iodide

Crystals (6809).

Fill another Direct
Reading Titrator

(0376) with Sodium

Thiosulfate Solution,

0.025N (4169).

Swirl to mix.
Solution will turn

yellowish brown.

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

Advertising