TEST PROCEDURE

PART 2 - ADDING THE REAGENTS

4

Be careful not to

introduce air into the

sample while adding

the reagents.

At this point

the sample has

For Kit Code 5860:
Add 8 drops of

*Sulfuric Acid, 1:1

(6141WT).

Cap and gently invert the bottle

to mix the contents until the

For Kit Code 7414:
Immediately use the 1.0 g

spoon (0697) to add one

level measure of *Sulfamic

Acid Powder (6286).

OR

NOTE:

2.

3.

Allow the precipitate

to settle below the

shoulder of the bottle.

5.

6.

NOTE:

precipitate and the reagent

have totally dissolved. The

solution will be clear yellow

to orange if the sample

contains dissolved oxygen.

been "fixed" and contact

between the sample and the

atmosphere will not affect the

test result. Samples may be

held at this point and titrated

later.

Immediately add

8 drops of

*Manganous

Sulfate Solution

(4167)
Add 8 drops of

*Alkaline

Potassium Iodide

Azide (7166).

AND

Remove the cap

from the bottle.

1.

4.

Cap the bottle and mix

by inverting several

times. A precipitate will

form.

Advertising