G, go to, Electroporation protocols decision tree, Initial experimental results – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual

Page 14

Advertising
background image

Once the best waveform is identified, use this tree to program the MXcell system to
improve other parameters.

Electroporation Protocols Decision Tree

Gene Pulser MXcell

Electroporation Guide

Exponential Waveform

Square Waveform

Optimize Capacitance

Manually program a capacitance

gradient (–10% increments)

around the value that yielded the

best results, keeping all other

parameters constant.

Optimize Voltage

Program a voltage gradient around

the value that yielded the best

results, keeping all other

parameters constant.

Optimize Voltage

Using the optimized capacitance

value, program a voltage gradient

(–10% increments) around the value

that yielded the best results,

keeping all other parameters constant.

Optimize Pulse Duration

Use the optimized voltage and keep all

other parameters constant. Vary pulse

values at  ms intervals around the

pulse that yielded the best

results (see protocols on page 10).

Optimize Resistance

Using the optimized capacitance and

voltage values, manually program a

resistance gradient (–10% increments)

around the values that yielded the

best results, keeping all other

parameters constant.

Optimize Number of Pulses

Use 2– shorter pulses that add

up to the pulse duration that yielded

the best results (see protocols

on page 10).

Initial Experimental Results

Visit us on the Web at discover.bio-rad.com

| Electroporation Decision Tree

14

Advertising