Section 5
Troubleshooting Guide
Problem
Cause
Solution
Background dark or uneven
Agitation too gentle.
Increase agitation speed during
gel incubation and ensure gel is
completely submerged.
Dodeca stainer or staining
Thoroughly clean Dodeca stainer
containers contain residue from
or containers with soap and
prior cleaning or staining runs.
water. Rinse only glass
containers with 50% nitric
acid, followed by thorough
water rinse to remove the acid.
Fingerprints or powder from
Use powder-free laboratory
gloves cause staining artifacts.
gloves.
Interfering components (glycine,
Increase fixing time or use
SDS, etc.) not completely
multiple fixing cycles to
removed from gel.
completely wash interfering
compounds from gel.
Sensitizing working solution
Prepare new sensitizing
improperly prepared.
solution with BRC.