Bio-Rad Trans-Blot® Cell User Manual

Page 5

Advertising
background image

1.2 Safety Instructions

Power to the Trans-Blot cell is supplied by an external DC voltage power supply. This

power supply must be ground isolated in such a way that the DC voltage output floats with
respect to ground. All of Bio-Rad’s power supplies meet this important safety requirement.
Regardless of which power supply is used, the maximum specified operating parameters for
the cell are:

300 VDC

Maximum voltage limit

200 Watts

Maximum power limit

50 °C

Maximum ambient temperature limit

Current to the cell, provided from the external power supply, enters the unit through the

lid assembly, providing a safety interlock to the user. Current to the the cell is broken when
the lid is removed. Do not attempt to circumvent this safety interlock, and always turn the
power supply off before removing the lid, or when working with the cell in any way.

Important

This Bio-Rad instrument is designed and certified to meet IEC 1010-1* safety standards.
Certified products are safe to use when operated in accordance with the instruction manual.
This instrument should not be modified or altered in any way. Alteration of this instrument
will:

Void the manufacturer's warranty

Void the IEC1010-1 safety certification

Create a potential safety hazard

Bio-Rad is not responsible for any injury or damage caused by the use of this instrument for
purposes other than for which it is intended or by modifications of the instrument not per-
formed by Bio-Rad or an authorized agent.

*IEC 1010-1 is an internationally accepted electrical safety standard for laboratory instruments.

2

!

!

!

Advertising