Appendix b. specifications – Bio-Rad Model 583 Gel Drying Supports User Manual

Page 13

Advertising
background image

APPENDIX B.

SPECIFICATIONS

Vacuum

Starting

> 29” Hg (23.4 Torr)

After 10 hours

> 28” Hg at 22° C (48.8 Torr)

At 35° C RT

> 26” Hg at start (99.6 Torr)

Flow rate

60 Hz

> 28 liters per minute

50 Hz

> 21 liters per minute

Fluid Capacity

Minimum (MIN)

4 liters

Maximum (MAX)

5 liters

Function Modes

Continuous or timed operation

Timer Control

0 to 3 hours, fully adjustable

Operating Conditions

Temperature

4 to 35° C

Humidity

10 to 90% relative humidity

Dimensions (D x W x H)

40 x 25 x 51.1 cm

Weight (dry)

14 kg (30 lbs)

Electrical connection

3-wire power cord.

Power requirements

115 V AC models

90-140 V AC, 50/60 Hz, 6 Amps

230 V AC models

200-264 V AC, 50/60 Hz, 4 Amps

Fuses required

Type T, 250 V AC

115 V AC models

8 Amp; 5 x 20 mm; slo blo (2)

230 V AC models

6.3 Amp; 5 x 20 mm; slo blo (2)

Regulatory

Certified to

I.E.C. 1010/Low Voltage Directive 73/23/EEC

Certified to

E.M.C. Directive 89/336/EEC (220 V only)/
EN55014 Emissions
EN50082-2 Immunity