Symptoms

Cause

Solutions

Sample wells cast

Comb should be placed 1 to 2 mm

through the gel. Sample

above the base of the running surface.

leaks along bottom of

running surface.

Sample overload

Dilute sample.

Bands sharp but too

Gel agarose percentage too high

Lower agarose percentage.

few bands seen

Incomplete digestion

Check enzyme activity, digest further.

High MW bands

Gel agarose percentage too low

Increase agarose percentage.

sharp; Low MW

Switch to polyacrylamide.

bands smeared

Gels crack

Too high voltage gradient,

Reduce voltage. Run gel at lower

especially with low melting

temperature.

temperature agarose or

low gel strength gels

Section 6
Product Information

6.1 Sub-Cell

®

GT Systems

Catalog
Number

Product Description

170-4401

Sub-Cell GT System, with 15 x 10 cm tray

170-4402

Sub-Cell GT System, with 15 x 15 cm tray

170-4403

Sub-Cell GT System, with 15 x 20 cm tray

170-4404

Sub-Cell GT System, with 15 x 25 cm tray

170-4481

Sub-Cell GT System, with 15 x 10 cm tray and gel caster

170-4482

Sub-Cell GT System, with 15 x 15 cm tray and gel caster

170-4483

Sub-Cell GT System, with 15 x 20 cm tray and gel caster

170-4484

Sub-Cell GT System, with 15 x 25 cm tray and gel caster

170-4405

Wide Mini-Sub Cell GT System, with 15 x 7 cm tray

170-4485

Wide Mini-Sub Cell GT System, with 15 x 7 cm tray and mini-gel caster

170-4406

Mini-Sub Cell GT System, with 7 x 7 cm tray

170-4486

Mini-Sub Cell GT System, with 7 x 7 cm tray and mini-gel caster

Sub-Cell Systems with PowerPac

Basic Power Supply, 100—120/220—240 V

164-0302

Sub-Cell GT Cell and PowerPac Basic Power Supply

164-0301

Wide Mini-Sub Cell GT Cell and PowerPac Basic Power Supply

164-0300

Mini-Sub Cell GT Cell and PowerPac Basic Power Supply

14

Advertising
This manual is related to the following products: