2 pulsed field conditions by dna size, 3 pulsed field conditions by organism – Bio-Rad CHEF-DR II System User Manual

Page 28

Advertising
background image

very large DNA (>3 mb) cannot be resolved on the gel and the field strength needs to be
reduced. Moreover, some large DNA molecules will not enter the gels at high field strengths.
Therefore, in selecting the field strength for an experiment, a compromise between run time
and resolution has to be made.

Electrophoresis Run Time

The electrophoresis run time is determined by the migration rates of the DNA molecules

under investigation. The migration rates, in turn, are affected by the switch time, field strength,
and field angle. As the migration rate of the DNA molecules decreases, the electrophoresis run
time needs to increase to adequately resolve the DNA molecules of interest.

5.2 Pulsed Field Conditions by DNA Size

In the table below are suggested run parameters that can be used for the various DNA

size ranges listed.

DNA

DNA

DNA

DNA

1-100kb

0.1 - 2.0 mb

2 - 4 mb

> 4 mb

% Agarose

1.0-1.2%

0.8-1.2 %

0.6-1 %

0.5-0.8 %

Buffer

0.5x TBE

0.5x TBE

1.0x TAE

1.0x TAE

Temperature 14° C

14° C

14° C

14° C

Voltage

6 V/cm

4.5-6 V/cm

2-3 V/cm

1.5-2.5 V/cm

Pulse Parameters

0.05-10 sec

10-200 sec

200-1800 sec

10-60 min

Run Times

2-15 hr

15-30 hr

24-72 hr

72-144 hr

5.3 Pulsed Field Conditions by Organism

In the table below are the run parameters for various types of DNA samples.

DNA

Agarose Switch time Run Time

DNA

size (kb)

Conc.

(seconds) (hours)Voltage Angle [Buffer]

Restriction

0.2-23

1.2%

0.01a

4

6 V/cm

120°

0.5x TBE

Fragments

5 kb Ladder

5-75

1.0%

1-6 b

11

6 V/cm

120°

0.5x TBE

Lambda Ladder 50-1,000

1.0%

50-90 c

22

6 V/cm

120°

0.5x TBE

Saccharomyces 200-2,200

1.0%

60-120 d

24

6 V/cm

120°

0.5x TBE

cerevisiae

Candida

1,000-4,000

0.8%

120 e

24

3.5 V/cm 106°

1.0x TAE

albicans

240

36

Schizo-

3,500-5,700

0.8%

1,800 f

72

2 V/cm

106°

1.0x TAE

saccharomyces
pombe

Dictostelium

3,600-9,000

0.8%

2,000-7,000 g

158

1.8 V/cm 120° 0.25x TBE

discodium

7,000-9,600

82

1.5 V/cm 120°

(a)

0.09 second single switch time for 3 hours.

(b)

Ramped switch time from 1 to 6 seconds over 11 hours.

(c)

Ramped switch time from 50 to 90 seconds over 22 hours.

(d)

Ramped switch time from 60 to 120 seconds over 24 hours.

(e)

120 secondswitch time for 24 hours followed by 240 second switch time for 36 hours.

(f)

30 minute single switch time for 72 hours.

(g)

Two blocks, with voltage change in the second block. Buffer temperature is 10 °C. Cox et al., Proc. Natl. Acad. Sci. USA, 87,