3 specifications – Bio-Rad CHEF-DR II System User Manual

Page 5

Advertising
background image

1.3 Specifications

These specifications pertain to the complete CHEF-DR II system. A system including

the Model 1000 Mini Chiller is also available.

CHEF-DR II Specifications

Drive Module

Dimensions

26 (depth) x 43 (width) x 14 (height) cm

Construction

Aluminum chassis

Weight

7.8 kg

Maximum voltage

250 VDC

Maximum current

500 mA

Electrode potentials

Dynamically regulated (feedback adjustment)

+/- 0.5%

Input voltage range

90-132 VAC/47-63 Hz/0.7 amps (maximum)
198-264 VAC/47-63 Hz/0.35 amps (maximum)

Fuses

0.5 amp Fast Blow for high voltage output (front panel)
1.0 amp Slow-Blow (100/120 V) or
0.5 amp Slow-Blow (220/240 V)

Environmental

Operating

50 °F (10 °C) to 90 °F (32 °C) temperature 30-80%
humidity

Storage

32 °F (0 °C) to 140 °F (60 °C) temperature 10-90%
humidity

Control Module

Dimensions

13.5 (depth) x 31 (width) x 17 (height) cm

Maximum current

500 mA

Allowable voltage gradients

0.6 - 6 V/cm, in 0.1 V/cm increments

Data entry

Keyboard

Power supply

250 V maximum, to allow maximum gradient of

6 V/cm,continuously adjustable

Functional

Switching range

0.1 sec to 65,000 sec (all electronic switching)

Maximum program blocks

2, with automatic execution

Maximum run time

999 hours per block

Input voltage range

90-132 VAC/47-63 Hz/1.7 amps (maximum)
198-264 VAC/47-63 Hz/0.85 amps (maximum)

Fuses

2.0 amp Slow-Blow (100/120 V) or
1.0 amp Slow-Blow (220/240 V)

Environmental

Operating

50 °F (10 °C) to 90 °F (32 °C) temperature 30-80%
humidity

Storage

32 °F (0 °C) to 140 °F (60 °C) temperature 10-90%
humidity

Electrophoresis cell

Dimensions

43 (depth) x 44 (width) x 11 (height) cm, horizontal format

Construction

Acrylic

Lid

Safety interlocked

Weight

8.5 kg

Electrodes

24, platinum (0.01 inch diameter)

Temperature monitoring

Via precision temperature probe mounted through lid
(optional)

2