| An Overview of Bio-Plex

Manager

MP

12

| Bio-Plex Manager MP Software

After the routine has been successfully run,

None is displayed indicating that

maintenance of the instrument is up to date.

Note:

There may be times when you have to run more than one recommended

routine before maintenance is up to date. For example, Enhanced Maintenance

may be displayed as the recommended routine if calibration or verification fails.

After the Enhanced Maintenance routine is successfully completed, Daily Start

Up is displayed in the Recommended Routine section. Run the recommended

routines until

None is displayed.

Tip:

If calibration, verification, or both routines fail after successive attempts,

contact Bio-Rad Technical Support for assistance.

None is displayed when maintenance is up to date

The maintenance of the MAGPIX instrument is up to date and you can now create

and run your protocols.