Cleaning the sample probe – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual
Page 36
Advertising
![background image](/manuals/600303/36/background.png)
4
| Maintaining the MAGPIX Instrument
34
| Bio-Plex Manager MP Software
Cleaning the Sample Probe
To clean the sample probe
1.
Turn off the MAGPIX instrument and unplug the power cord.
2.
Remove the sample probe.
3.
Use either or both methods to clean the sample probe:
Bath sonicator — place the tip of the sample probe in the bath sonicator
for 2–5 min.
10 ml syringe — force distilled water through the tip of the sample probe to
its large end.
4.
Replace the sample probe. Screw the black fitting on top of the sample probe
until it clicks.
5.
From the Maintenance view, run the Adjust Probe Height procedure.
Adjust Probe Height