Getting Started Guide

| 33

4

Maintaining the MAGPIX
Instrument

This chapter describes the periodic maintenance that must be performed on the

MAGPIX instrument to keep it working according to the manufacturer’s

specifications.

Daily Maintenance

Each day you use the MAGPIX instrument, perform these maintenance tasks:

At the beginning of the day at the beginning of each day you use the

instrument, run the recommended routine. Check the Recommended

Routine section in the Maintenance view. On most days this is the Daily

Start Up routine.

At the end of the day — at the end of each day you use the instrument,

run the Shut Down routine.

Weekly Maintenance

Clean the sample probe once a week to keep it free of any debris. Whenever you

remove the sample probe from the probe assembly, you must recalibrate the probe

before you read another plate. Therefore, cleaning the sample probe must always

be followed up with running the Adjust Probe Height maintenance routine.