background image

Differences Between Bio-Plex

Manager

MP and xPONENT

Getting Started Guide

| 7

Differences Between Bio-Plex

Manager

MP and

xPONENT

If you have used xPONENT software with MAGPIX, there are some differences in

how Bio-Plex

Manager

MP software works that you should be aware of:

Adjusting probe height — adjusting the probe height is simple and

reproducible. The software uses the probe height adjustment plate to set

the probe height for the two locations in the off-plate reagent area and four

heights for consumables. These settings cover most consumables. The

height used in the four settings is indicated on the probe height adjustment

plate and in the Bio-Plex

Manager

MP User Guide.

Start-of-day maintenance — run the Daily Start Up routine at the

beginning of each day. With Bio-Plex

Manager

MP, you do not have to

keep track of whether the instrument requires calibration and verification or

only verification on any given day. The software ensures that the correct

routines are run each day. The Daily Start Up routine also runs additional

fluidics routines if the instrument has been idle for some time.

Recommended routines — the software always recommends a routine if

it is required for the proper running of the instrument. There may be times

when a recommended routine requires additional steps; follow these steps.