Run assay, Run assay 17, Considerations – Bio-Rad Bio-Plex Pro™ Human Isotyping Assays User Manual

Page 19: Considerations when using a vacuum manifold

Advertising
background image

8. Run Assay

Considerations

n

Bring all buffers, diluents, diluted standards, diluted coupled beads, and

samples to room temperature before use

n

Use calibrated pipets and pipet carefully, avoiding bubbles. Use new

pipet tips for every volume transfer

n

Pay close attention to vortexing, shaking, and incubation instructions.

Deviation from the protocol may result in low assay signal and
assay variability

n

Assay incubations are carried out in the dark at 850 ± 50 rpm. Cover

the plate with sealing tape and protect from light with aluminum foil

Table 7. Summary of wash steps and incubations. After each assay step, select the

appropriate Bio-Plex Pro

wash station program or perform the appropriate manual wash step

as summarized below.

Considerations When Using a Vacuum Manifold

n

After each incubation, place the filter plate on a calibrated vacuum

apparatus and remove the liquid by vacuum filtration

n

To wash, add 100 μl wash buffer to each well and remove the liquid as

before. Ensure that all wells are exposed to the vacuum

n

Thoroughly blot the bottom of the filter plate with a clean paper towel

between each vacuum step to prevent cross contamination

n

Place the assay plate on the plastic plate holder/tray as needed

n

Before each incubation, gently cover the plate with a new sheet of sealing

tape. Avoid pressing down over the wells to prevent leaking from the bottom

17

Bio-Plex Pro or

Bio-Plex Pro II

Handheld Magnet or

Pro II Wash Station

Wash Station

Vacuum Manifold

Assay Step

Magnetic Program

Vacuum Program

Manual Wash Steps

Add beads to plate

MAG x2

VAC x2

2 x 100 μl

Sample incubation
Detection Ab incubation

MAG x3

VAC x3

3 x 100 μl

SA-PE incubation

Advertising