Electrode Cleaning
A dirty pH electrode can lead to erroneous pH readings and shorten the lifetime of the electrode. Most contami-
nant’s will be either proteins that have adsorbed onto the glass membrane or buffer and salt crystals that have
blocked the pH probe pores.

Removal of protein deposits: Immerse the electrode in a 1% solution of proteases, such as trypsin or pepsin, in
0.1M HCl for no more than 5 minutes. Rinse with deionized water to remove all traces of HCl.

Removal of salt deposits: Immerse the electrode in 0.1M HCl for 5 minutes followed by 0.1M NaOH for 5 minutes
and then 0.1M HCl for 5 minutes. Rinse the electrode with deionized water to remove all traces of HCl.

Storing Your pH Electrode Between Uses
When pH readings are made infrequently (i.e., several days or weeks apart), the electrode should be stored in its
plastic bulb filled with electrode storage solution (available from VWR or Orion) or pH 4 buffer in 3.5 M potassium
chloride.
Do Not Allow the Electrode to Dry Out.

Trouble Shooting

Symptom

Possible Cause

Possible Solution

Noisy or unstable Readings

Air bubble on membrane

Remove air bubble

Drift

Solution temperature is changing

Allow all solutions to come to thermal equilibrium.

Leakage around pH probe

Loose nut

Tighten pH electrode nut

Leakage around pH probe

Missing or damaged o-rings

Replace o-rings

Leakage around fittings

Loose fitting

Tighten fitting

Leakage around fittings

Cracked or loose ferrule

Replace fitting

Technical Assistance

For additional help, contact your local Bio-Rad representative. In the United States, call Chromatography
Technical Support at 1-800-4BIORAD.

Ordering Information

760-2040

DuoFlow pH Monitor

760-2042

pH Electrode

760-2044

Flow Cell

750-0502

Signal Import Module

910-1828

o-ring, pH flow cell, size 2-009

910-3014

o-ring, pH electrode, size 2-018

760-2046

BioLogic pH Monitor tubing kit – for connection between Conductivity flow cell and pH flow cell inlet
includes
1 x 28 cm 1/16” OD x 0.02” ID orange PEEK, rated to 5000 psi
1 x 28 cm 1/16” OD x 0.03” ID green PEEK, rated to 3000 psi