Data view version for the system – Bio-Rad LP Data View Software for BioLogic LP System User Manual

Page 4

Advertising
background image

Data View Version

for the

system

Every effort has been made to ensure that the

Data View software is as trouble-free as

However, any

software application program, certain “bugs” are found after the

is released. This document lists

known

bugs

in

version

and ways to work around them. These bugs will be fixed in a future release of

View. please monitor our

web site at

for information concerning software upgrades.

Bug

Work-around

International versions of

The

Data View program does not

Windows’

work correctly with international
versions of Windows where the comma

and the period characters used in

formatting numbers are reversed from
the US version of Window?.

Manual mode fraction

collection

Time Axis Window Text Box

The

Data View program does not

automatically record data when
fraction collection is started in manual
mode of the

system and

does not reset the fraction number
counter correctly in manual mode.

If an incorrect entry, for example a

blank entry or a decimal point, is made
in the Time Axis Window Text Box
and the user clicks on “Set” or presses

the “‘Enter” key, then the program will
crash display the message “Run time
error ‘

Type mismatch” and crash.

Any unsaved data will be lost, but if
Data Logging was ON the Log file can
be used to recover the data when the
program is started again. To prevent
this problem, always ensure that the

correct numerical entry is made
beforehand.

Change the number formatting charac-
ters to match the US convention. This
is done by opening the “Regional
Settings” program within the Windows

“Control Panel”. Select the

“Number” Tab and change the
“Decimal symbol” to a period and
change the “Digit grouping symbol” to
a comma

Restart your computer

after making these changes. If this
does not work for any reason, you
must install the US version of

Windows@

Click the “record” button on the
bar of

Data View to record data in

manual mode before starting fraction
collection and click on the “stop” but-
ton when fraction collection has been
stopped. To reset the fraction number

counter without running a method on
the

you must exit Data

View

and

then

start

the

program again.

If this situation arises, then PAUSE the
run from the

Controller.

the

Data View program,

save any log file which will contain the

run data and click on

“record”. Then resume the run from

the

Controller.