Saving a Log file as a Data file

If a Log file is saved as a Data file

To prevent the problem, enter a different

’ .

without entering a different name for
the tile, the original. Log file remains

name when saving a Log file as a Data
file.

but the Data file is not saved. Data
View also incorrectly presents a
warning about “replacing an existing
file”. When this problem occurs, no
data is lost.

Setting axes

Viewing multiple open

applications

If multiple entries are made in the Axes

the ‘Set” button as appropriate

Settings Text Boxes for different axes

for each axis.

before clicking on the “Set!’ button or
pressing the “Enter” key, you must
click on each “Set” button for all
entries to take effect. The “Enter” key
will only activate the “Set” button for
the axis Text Box which has the focus.

When data recording is started

The user should click on the program

automatically as the

they were using previously or use

System starts a run, Data View takes

to send the focus back to the

the computer focus away from other

desired program.

applications and does not return it even
if

Data View is minimized.

Scroll bars

While recording -data and using

The solution is to use the Full View

Window mode for the Time Axis, the

setting for the Time Axis to see all data

scroll bar is not active.

and then use the mouse again to “zoom
in” on a new region.

Viewing runs of less than one

minute duration

Data View does not use the scroll

The Full View setting will show all

bar correctly when using the Window

data and the mouse can be used to

setting for the Time Axis when less

rubberband the data if necessary.

than 1 minute of data is present.

Printing in Landscape format

The program does not print in

No work-around for this release.

Landscape mode in Windows NT.