2 Bio-Scale MT Column Instruction Manual

1.3 Column Specifications

All wetted parts are bio-compatible (Tefzel, polypropylene,
polyphenyl sulfide, borosilicate glass and buna-N rubber) and
resistant to cleaning and sanitation solvents including 1.0 M NaOH
and 25% acetic acid. The column can be used with most common
aqueous buffers and certain organic solvents including alcohols
and acetonitrile. Do not use halogenated hydrocarbons, aromatic
solvents, or tetrahydrofuran (THF). A detailed solvent compatibility
table is given in Section 4.

Table 1. Column Characteristics.

MT2

MT5

MT10

MT20

Column
Volume, ml

1.9 – 2.3

4.6 – 5.7

9.5 – 11.3

19.4 – 21.9

Column
Dimension, mm

7 x 50 – 60

10 x 59 – 72

12 x 84 – 100

15 x 110 – 124

Maximum
Operating
Pressure, psi

1, 000

750

600

500

1145 psi = 1 MPa. 14.5 psi = 1 bar. Packing material may not withstand high
pressures. See packing material manufacturer’s specifications.