Bio-Rad UNOsphere™ Rapid S Media User Manual

Page 23

Advertising
background image

Section 5
Cleaning-in-Place (CIP) and
Sanitation

If a column no longer yields reproducible results, the
media may require thorough CIP and sanitation to
remove strongly bound contaminants. Acceptable
CIP agents include 25% acetic acid, 8 M urea, 1%
Triton X-100, 6 M potassium thiocyanate, 70%
ethanol, 30% isopropyl alcohol, 1 N HCl, 1 N
NaOH, and 6 M guanidine hydrochloride.

1. Sanitize the support in the column with 2–4 bed

volumes of 1.0 M NaOH at 50–100 cm/hr while
maintaining a minimum contact time of 40 min.

2. To reequilibrate the column, wash the column

with 2–4 bed volumes of 0.5–2 M NaCl solution
(containing 50–100 mM buffer).