Bio-Rad Aurum™ Total RNA Mini Kit User Manual
Page 23

Section 9
Troubleshooting Guide
Problem
Possible Cause
Recommended Solution
Genomic DNA
Incomplete DNase I
Increase DNase I digest
contamination
digest
time
Inactive DNase I
Store reconstituted
DNase I in a nonfrost-
free freezer. Avoid
freeze-thaw cycles.
Aliquot reconstituted
DNase I for single use
only
Incorrect preparation
Use only the DNase
of DNase dilution
dilution solution
provided in the kit to
dilute the DNase
Excessive amount of
Reduce amount of
starting material
starting material used
RNA degradation
RNase contamination of
DEPC-treat all hand-
user-made solutions
made solutions.
and/or plasticware
Decontaminate all work
surfaces. See Section
5 for more details
Endogenous RNases
Work quickly through
the steps prior to the
addition of lysis solution
Clogging of RNA
Excessive amount of
Reduce amount of
binding column
starting material
starting material used
Poor disruption and/or
Increase intensity/
homogenization
duration of disruption
and/or homogenization
Switch to more intense
disruption and/or
homogenization method
Incomplete digest with
Increase duration of
lysozyme or lyticase
lysozyme or lyticase
digest. Use fresh
enzyme
19
4110133B:4110133A.qxd
1/16/2009
2:35 PM
Page 19