Bio-Rad Aurum™ Total RNA Mini Kit User Manual
Page 24

20
Problem
Possible Cause
Recommended Solution
Low eluate volume
Insufficient centrifugation
Add 1–3 min to the
(<60 µl)
time during elution
centrifugation time during
elution
Column clogging
See problem “Clogging
of RNA binding column”
High eluate volume
Low stringency wash
Add 1–3 min to the
(>80 µl)
carryover in eluate
centrifugation time
after the final wash step
Low RNA yield
Low amount of
Increase starting
starting material
material amount up to
the maximum indicated
for the specific starting
material type
Excessive amount of
Reduce amount of
starting material
starting material used
Poor disruption and/or
Increase intensity/
homogenization
duration of disruption
and/or homogenization
Switch to more intense
disruption and/or
homogenization method
Incorrect use of wash
Add the appropriate
solutions
volume of 95–100%
ethanol to the wash
solutions before initial
use
Incorrect preparation of
Use only the DNase
DNase dilution
dilution solution
provided in the kit
to dilute the DNase
Low sample eluate
See problem “Low
volume
eluate volumes (<60 µl)”
Inefficient elution
Preheat the elution
solution to 70°C in
water bath prior to
the elution step