13

Problem

Possible Cause

Possible Solution

More than one band

Presence of multimers –

Try different host or

on analytical gel

typical and variable

growth conditions

depending upon plasmid,
bacterial host, etc.

Nicking due to

Use endA- host such as

endonuclease activity

JM109, DH5

a, or

XL1-Blue

Prolonged exposure to

Add neutralization solution

alkaline conditions

within 5 min after
addition of lysis solution

Plasmid nicked/

Use of endA+ bacterial

Use endA- host such as

degraded

host

JM109, DH5

a, or

XL1-Blue

Plasmid prep

Plasmid nicked, degraded Check integrity of plasmid

performs poorly in

or denatured

on an analytical gel

enzymatic reactions

Use endA- host

Section 8

Ordering Information

Catalog #

Description

732-6470

Aurum Vacuum Manifold, includes column adaptor plate,
4 replacement luer caps, A and B stages, waste collection
tray, vacuum regulator and gauge, tubing, protocol overview,
and instruction manual

732-6400

Aurum Plasmid Mini Kit, 100 preps, includes 100
plasmid binding columns, 100 capless collection tubes (2.0
ml), reagents, protocol overview, and instruction manual

Advertising