Flow Properties

Fig. 1. Nuvia cPrime media pressure vs. flow performance for a
20 cm diameter column and a 20 cm bed height; compression ratio

1.17. Nuvia cPrime media have fast mass transfer properties allowing

users to achieve high productivity at fast flow rates. The media should

be run at the highest linear velocities that allow good separation and are

allowed by the column and chromatography system specifications. A

linear flow rate of 300 cm/hr and a 20 cm bed is a recommended starting

point.

Fig. 2. Effect of flow rate on Nuvia cPrime media binding capacity for

lactoferrin.

Column dimension: 1.1 x 9.6 cm
Sample: 5.25 mg/ml lactoferrin

Nuvia cPrime Media vs. Pressure Flow