Model 6712 Point-of-Sale Interface Scale User’s Manual

14

To abort the CAL
Mode at any time,
press the SW-1
switch.

Step-by-Step
Instructions for CAL
Mode

Calibration Mode

Calibration (CAL) Mode

Calibrating your
scale requires a
certified test weight
to ensure accurate
weighing.

The calibration (CAL) mode menu lets you
calibrate your scale. The items in the calibration
menu are as follows:

POUNDS/KILOGRAMS/OUNCES – Select the
unit of measure (lb, kg or oz).

CAPACITY – Select the capacity / resolution of
the scale.

Follow these steps to calibrate your scale. Refer
to Figure 2.

1. From the normal weighing mode, press SW-

1…

DIAG is displayed.

2. Press the ZERO key until…

CAL is displayed.

3. Press the TEST key to start calibration…

OUNCE is displayed.

4. If you want to change the unit of measure,

press the ZERO key to toggle between
OUNCE, LB or KILO. When the choice you
want is displayed, press the TEST key…

The current capacity selection is then
displayed.