2.0 GETTING STARTED

SPECIFICATIONS

Sensors:

Concentric plate displacement current electric

field sensor.

8 inch diameter magnetic field sensing loop

Switch selectable between electric and

magnetic fields

Sensitivity:

Electric fields - 1 - 1999 V/m

Magnetic fields - 1 - 1999 mA/m

Features:

All function and range selection changes are
made by use of the membrane switch
keypads on the front panel.

Internal electronic range switching allows full
auto-ranging on either the Electric or Magnetic
field functions.

Max-hold feature stores and displays highest
reading

Response:

True RMS field measurement for accurate
measurement of non-sinusoidal waveforms

Frequency
Response:

Electric fields:

±0.5 dB, 10 kHz to 100 kHz

+0.5/-2 dB, 2 kHz to 300 kHz

Magnetic fields:

±0.5 dB, 12 kHz to 200 kHz

+0.5/-2dB, 8 kHz to 300 kHz

Determined with applied electrical field
strength of 143 V/m and an applied magnetic
field strength of 166 mA/m in auto-range
mode.

Temperature
Range:

20°F - 105°F
(-10°C - 40°C)

Power:

Two (2) nine-volt alkaline batteries (NEDA
1604A, Duracell MN1604, or equal)

Output:

Liquid crystal display, preamplifier output via
phono jack (analog signal from
sensor/preamplifier equal to 1 mV/(mA/m),
digital fiber optic signal (for remote reading via
connection to HI-3616 Fiber Optic Remote
Control)

Advertising