ETS-Lindgren HI-3603 VLF Survey Meter User Manual

Page 38

Advertising
background image

HI-3603 VDT/VLF Survey Meter

© ETS-Lindgren, August, 2005

Revision E, Part # H-600042

38

Standard/Reference Frequency

(kHz)

E (V/m)

RMS

H (gauss)

RMS

ACGIH

0 - 0.1

25000

---

0.1 - 4

2500/f

1

0.6/f

1

4 - 30

625

---

IEEE C95.1-1991

3 - 100

614

2.05

IRPA
(gen. pop. 24 hr.)

50/60 (Hz)

5000

1.0

Swedish
Guidelines

2

0.005 - 2

25

0.0025

2 - 400

2.5

0.00025

UK
(1986) (occ)

750 Hz -
50 kHz

2,000

1.25

UK
(1986) (public)

750 Hz -
50 kHz

800

0.05

USSR
(public) (Slesin,
1985)

0.03 - 0.3

25

---

(1)

Frequency in kHz

(2)

These guidelines are not based on biological effects.

They are based on what was technically possible.

Table 2: Radiofrequency Exposure/Emission Standards

Pertinent to the VDT Frequency Range

SPECIALIZED MEASUREMENTS WITH THE HI-
3603

This section is intended for those individuals wishing to obtain
more detailed information on the characteristics of VDT
emissions than might normally be considered routine.

FLYBACK FREQUENCY MEASUREMENTS

The HI-3603 is equipped with a waveform output jack located on
the bottom of the case which provides a signal from the
preamplifier in the input module. This is an analog signal that has
essentially the same shape as that of the electric or magnetic
field waveform that is being measured. Connection of a

Advertising