DISP 3/4 - Pressing the DISP pad will shift the display from 4 to
3 digits and back again. In some situations where the measured
field is fluctuating, the three digit display will make reading the
instrument easier.

CLEAR DATA - Clears values stored in the data logging
memory. Pressing this keypad will clear all data stored in the
data logging memory. When the keypad is pressed a flashing clr
is indicated. Holding the keypad until 000 is shown will clear the
memory.

LOG - Pressing the LOG pad will store the current reading in the
data logging memory. When the LOG pad is pressed, the
identification number of the value (1-112) is displayed for about
one second followed by the stored value. Immediately upon
releasing the key, a new value may be logged by again pressing
the LOG pad. Up to 112 values can be stored in this memory.
When the memory is filled, successive operations of the LOG
key will store the most current reading in memory location 112.

PREV - Pressing the PREV key displays the last stored data
value. When the key is pressed, the identification number of the
value is first displayed for about one second followed by the
stored value. The stored value is displayed as long as the key is
depressed. About two seconds after releasing the key, the
display returns to the current reading. Successive operations of
the key will move the displayed value down towards the
beginning of the memory (value identification #1). If the PREV
key is operated while viewing value #1, the identification value
will "wrap around" to the highest stored identification number.

NEXT - Pressing the NEXT key displays the next value in the
data logging memory. If the NEXT key is operated while viewing
the last stored value, the identification number will "wrap around"
to value number one.