Measured GTEM! Performance

|

31

5.0 Measured GTEM! Performance

Several different measurable quantities exist that illustrate the performance of a
Gigahertz Transverse Electromagnetic (GTEM!™) Cell. Typical measurements
are discussed in this section.

GTEM! Performance Qualification – VSWR

The Voltage Standing Wave Radio (VSWR) of a terminated transmission line is a
fundamental performance parameter. The value of the VSWR reading is a
measure of mismatch; therefore, it completely defines the capability of a
GTEM! to transfer power to or from interconnected 50-ohm RF test equipment.
The VSWR is a qualifying measurement for GTEM! cells. Most GTEM! cells are
pre-assembled at the factory and the VSWR verified at that time. Following is the
VSWR of a typical Model 5407 measured with an HP8753C vector network
analyzer.

T

YPICAL

VSWR

M

EASUREMENT

M

ODEL

5407

GTEM! Electromagnetic Field Response

E-F

IELD

F

REQUENCY

R

ESPONSE

The E-field frequency response is defined as the measurement of the electric
field strength at a specific point in the GTEM! measured as a function of
frequency and normalized to a constant forward input power.