GTEM! Use

5. Rotate the EUT through further 90-degree rotations to the right, such that

the y-axis is replaced by the z-axis, and so on, as shown in Typical EUT
System Installed in Large GTEM! (Additional 90

°

Rotation) on page 45.

6. Measure the RF voltages emanating from the EUT as a function of

frequency over the desired frequency range. These measurements are
referred to as V

zxy

versus frequency.

The detector selected (peak, quasi peak, or average) must be used for all
three-position measurements and the frequency of the measurements must be
the same for all positions.

After completion of these measurements, the tester should have a matrix of
measurements that consists of a frequency and three associated RF voltage
measurements all made with the same detector function. A value for the noise
measured at a given frequency where signal components were measured may
be necessary to complete the measurement set.

To assist with the computation, you may download an MS-DOS®-based utility
program from the ETS-Lindgren website at

www.ets-lindgren.com

. This type of

program should be used because the computations are too complex to be
performed by hand in a reasonable amount of time.