9 version menu, Version menu – Rice Lake 320IS Plus Intrinsically Safe Digital Weight Indicator - Installation Manual User Manual
Page 46

40
320IS Plus Installation Manual
3.2.9
Version Menu
The VERS menu is used to check the software or hardware version installed in the indicator. There are no
parameters associated with the Version menu; when selected, the indicator displays the installed software or
hardware version number.
Figure 3-14. Version Menu
TWZERO
65300
number
Tweak zero. Adjust the analog output zero calibration. Use a multimeter to monitor the analog
output value. Press and hold
or
to adjust the output. Press
to save the new value.
Default value becomes 11000 if OFFSET is set to 20%.
TWSPAN
65300
number
Tweak span. Adjust the analog output span calibration. Use a multimeter to monitor the analog
output value. Press and hold
or
to adjust the output. Press
to save the new value.
Default value becomes 54900 if OFFSET is set to 20%.
ALG OUT Menu
Parameter
Choices
Description
Table 3-9. Analog Output Menu Parameters (Continued)
Note
Note
SOFTWR
Software Version
I/O Module Version
XXXXXXX
XXXXXXX
XXXXXXX
XXXXXXX
VERS
ALGOUT
DIG IN
PROGRM PFORMT
SERIAL
CALIBR
CONFIG FORMAT
XXXXXXX
SETPNT
I/OMOD
Displays “No Ver”
if no I/O Module is
connected