3-1

CHAPTER 3 REPAIR PROCEDURES

3.1 REPAIR PROCEDURES

This section of the manual contains detailed disassembly procedures of the balance. Refer to Figure

5-1. Before disassembling the balance, it should be noted that components inside of the balance are

delicate and need to be handled with care. It is imperative that the Transducer (8) should never be

subjected to any excessive torque, stress, or abrasion as damage may result. Once the balance has

been disassembled, any small scratch or abrasion made to any of the Individual Strain Gauges will

render the entire Transducer (8) unusable. The Printed Circuit Board (9) contains integrated circuits

which employ CMOS technology, therefore, caution must be exercised so as not to subject any of these

components to static electricity discharge. When servicing, a wrist Ground Strap with a 10 Megohm

series resistor to earth ground is highly recommended. The Printed Circuit Board (9) should be

handled by grasping the edges only and never placing fingers on any of the runs or traces.

3.1.1 Cover Removal

To disassemble the balance, proceed as follows:

1. Turn the balance off and if using an AC Adapter, disconnect it from the balance.

2. Remove the Pan (1, 2 or 3) from the balance. Pan sizes and shapes vary.

3. Place the Shipping Lock in the locked position.

See Figure 3-1. This will help protect the

Transducer.

12 Battery Compartment Cover

Cover Screw

Cover Screws

4. If a battery is being used, turn the balance over and remove the Battery Compartment Cover

(12) as shown in Figure 3-2.

Figure 3-1. Shipping Lock.

Figure 3-2. Bottom of Balance.

Locked

Position

Pointer

Unlocked

Position