Rice Lake Scout Pro Series Portable, Ohaus - Service Manual User Manual
Page 34
A-6
APPENDIX A SERVICE TOOL INSTRUCTIONS
A.6
USING CONFIG TOOL AFTER REPLACING THE MAIN PC BOARD (CONT.)
5.
The balance displays Err 9.
6.
Start the ConfigTool software, see A.5
7.
An error message will be displayed User Enable Units & User Variable Check Sum
Verify Error, click on OK.
8.
Set RS232 to the correct com port, Baudrate = 2400, ByteSize = 7, Parity = None.
9.
The balance will respond to the V command but not the others.
10. The bar graph will respond to pressure on the platform.
11. Enter the balance serial number.
12. Click on OK.
13. Click on READ.
14. The balance should read the correct model and serial number.
15. Perform the factory calibration. See B.2
16. The unit is finished. Remove the RS232 option or the jumper on the customers RS232
option.