Computer Interface

Welch Allyn TM286 Auto Tymp

During Printing

NO DATA AVAILABLE

This message appears if the

button is pressed and no results are stored.

NOT AVAILABLE

This message appears if the

button is pressed and the computer is not properly

connected.

Data Transfer Program Mode

The Data Transfer Program mode is used to modify the TM286 USB interface
configuration parameters to match the computer's USB Port settings.

Enter the Program mode by selecting the Program (PROG) button. Move the cursor to
Data Xfer Config and press the

button to enter the submenu.

The following screen appears the first time the Data Transfer Program mode is entered
showing the factory default settings.

These selections fall into three groups:

Baud rate

Parity and data bits

Flow control

The default setting for each group has an asterisk (*) before it so that it is easy to scan the
settings for each group.

Selecting default settings for any of the groups is achieved in the same manner as the
Program mode. Use the

or

buttons to move the solid square cursor down or up

to the setting that you wish to select. Press the button. The word SAVED will appear in
the lower right corner of the screen and the asterisk (*) will appear in front of the new
setting.

To exit the Data Transfer Program submenu, move the cursor to the

press the

button. This will return to the Program mode which can be exited by selecting Program
(PROG).

PROGRAM MODE - DATA TRANSFER

* 115.2 kBAUD

* NO PARITY + 8-BIT

57.6 kBAUD

ODD PARITY + 7-BIT

38.4 kBAUD

EVN PARITY + 7-BIT

19.2 kBAUD

SPC PARITY + 7-BIT

9600 BAUD

* XON/XOFF DISABLED

4800 BAUD

XON/XOFF ENABLED