Standards, specifications, and accessories, Standards and compliance, Tympanometry modes – Welch Allyn TM286 Auto Tymp - User Manual User Manual

Page 65: 8 - standards, specifications, and accessories, Standards and compliance tympanometry modes

Advertising
background image

8

61

Standards, Specifications, and
Accessories

Standards and Compliance

The TM286 meets

Tympanometry modes

Standards:

UL 60601-1 Medical Electrical Equipment Requirements for Safety

IEC/EN 60601-1 General Requirements for Safety

CSA C22.2 No.601-1-M90

ANSI S3.39-1987 Aural Acoustic Impedance Admittance (Type 3)

IEC 60645-5 Aural Acoustic Impedance/ Admittance (Type 2)

ANSI S3.6-2004 Audiometers (Type 4)

IEC 60645-1 Pure Tone Audiometers (Type 4)

GL2005-00014 (ASHA 2005) Guidelines for Manual Pure-Tone

Threshold Audiometry

Protective Classification:

This system is intended for continuous operation and has a protective

classification of Class 1.

Class 1 Type B

Probe Tone:

226 Hz, ±2%

Sound Pressure Level:

226 Hz: 85.5 dB SPL, ±2.0 dB, measured in a 2.0 cm

3

coupler

Harmonic Distortion:

<3%

Admittance (Compliance) Range:

226 Hz: 0.0 to 1.5 cm

3

or 0.0 to 3.0 cm

3

Notes:
1. The range is automatically selected based upon the amplitude of the
compensated tympanogram.
2. The maximum uncompensated (ECV + tympanogram peak) admittance
(compliance) range is 0 to 5.0 cm

3

.

3. ECV/cavity limits for initiating pressurization is 0.2 to 5.0 cm

3

.

Compliance Accuracy: ±0.1 cm

3

or ±5%, whichever is greater.