Welch Allyn Atlas Monitor - User Manual User Manual
Page 32

20
Atlas Monitor showing Impedance Respiration waveform display,
controls, and ECG connector
Setting alarm limits –To set the Respiration
alarm limits, use the
Select
and
Set
buttons on
the left side of the monitor to follow these steps:
• Press the left Select button to cycle through
MAP
HI
and
LO
,
Heart Rate
HI
and
LO
,
Respiration
HI
and
LO
, then (model 623xx
only) CO
2
HI
and
LO
.
• Each push of Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and
HI
or
LO
will flash,
indicating which limit is being changed.
• Press the Set button up or down to raise or
lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).
• Press the Select button again to go to the
next limit, or press it several times until
none of the measurements flash and no
HI
or
LO
indicators are lit. The instrument is now
in its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).