A.9 I/O Communications (Models 622xx & 623xx)

Connector

RJ45 socket on rear of case

Nurse Call Signal

Normally open relay contacts between pins #1 and
#8 of RJ45 connector. Contacts closed when any
alarm is audible.
Normally closed relay contacts between pins #2
and #8 of RJ45 connector. Contacts open when any
alarm is audible. Contacts rated for 1 A, 240 VAC.

Serial Interface

Bi-directional RS232; 9600 baud, 8 bit, 1 stop bit,
no parity, no flow control

Protocols

Pangea SIO command interface
Welch Allyn serial communications protocol



A.10 End Tidal CO

2

(Model 623xx)

Measurement Method

Sidestream

Measurement Range

0 to 98 mmHg
Percentage and kPa conversions are based on
current barometric pressure.

Measurement Accuracy

Ambient temperature 15ºC to 45ºC:
0 to 40 mmHg: ±3 mmHg;
41 to76mmHg: ±8% of reading;
77 to 99 mmHg: ±10% of reading.

Ambient temperature 5ºC to 15ºC and 45ºC to
55ºC:
0 to 40 mmHg: ±4 mmHg;
41 to76mmHg: ±10% of reading;
77 to 99 mmHg: ±12% of reading.

Accuracy unspecified for breath rates >30
breaths/minute.

Standard Conditions

Automatic compensation to comply with BTPS.
BTPS CO

2

= (internal measurement) x 1.03

Breath Rate Range

5 to 100 breaths/minute

Breath Rate Accuracy

±1 breath/minute or ±5%, whichever is greater

Display Units

mmHg, %, kPa; user selectable

Resolution 1

mmHg

Alarms

High and Low CO

2

measurement

High and Low Respiration Rate