Technical alarms

Technical alarms occur when a problem occurs with such things as the patient device,

the patient location or ID, an accessory connected to the patient device, or the network.

1.

System message area. If multiple alarms are active, click

to cycle through the messages for each active

alarm.

2.

Patient tile. If a patient tile is in a technical alarm state, the border around the tile changes to a color relative to

the level of the alarm. No colored borders appear around the patient's parameters.

Main Monitoring screen notifications

Notification

Description

System message area

The area changes color and displays a message with an accompanying status icon or

button.
If multiple alarms and information messages are active, the System message area shows

the highest priority alarm. If the alarms are equal in priority, the most recent alarm

message appears. You can cycle through the messages for each active alarm.

Patient tile

The border color changes. Click the alarming tile to pause an alarm audio tone. Visual

indicators persist during an audio-paused condition.